Open the packing box and check whether the joints on the vacuum glove box are loose during transportation. If there is any looseness, please tighten the corresponding screws, and then you can try to vacuum.

The method of vacuuming and inflating:

1. This product is divided into two parts: the main chamber and the antechambers. The main chamber cannot be evacuated separately, and the antechambers can be evacuated separately.

2. First open the door inside the antechamber, and then close all the glove interface glands, valves and doors outside the antechamber.

3. Open the three-way valve between the glove ports (rotate counterclockwise), connect the space between the glove interface gland and the glove to the chamber, so that the inside and outside of the glove (box and interface) are simultaneously evacuated and maintained . The air pressure is balanced on both sides.

4. Connect the vacuum pump to a valve in the antechamber, turn on the vacuum pump, and slowly open the valve to pump air to the system (if the valve is opened too fast, it may cause the gloves to expand).

5. When the pointer of the vacuum gauge drops and stabilizes at -0.1MPa, the vacuum is completed. At this time, the valve should be closed first and then the vacuum pump.

6. Then fill the chamber with inert gas through another valve. The vertical pressure is slightly higher than the atmospheric pressure (barometer index is 0MPa), so that the pressure inside and outside the chamber is basically balanced, and the inside is slightly higher than the outside. Ensure that the atmosphere does not penetrate into the inside of the chamber(if you want to make the gas in the chamber more pure, it is best to repeat the above extraction and inflation process several times). Turn off the three-way valve connected to the glove interface, open the gland on the glove interface, and you can operate.


Second, the method of putting objects in and out of the main chamber:

1. How to put the object into the main chamber: After you have completed the operation of "(1) Vacuum and inflate" above, if you need to put the object into the main chamber, please close the inner door of the antechamber and open it. The outer door of the antechamber, put the objects in, then close the outer door, and use the two valves on the antechamber to pump and inflate. Make sure that the pressure in the main chamber and the antechamber is basically balanced with the atmospheric pressure before you can open the door inside and take the objects into the main chamber.


2. The method of taking out the objects in the main chamber: Put the objects in the antechamber first, and close the antechamber door tightly. If the antechamber indoor air pressure is basically balanced with the atmospheric pressure, the antechamber outdoor door can be opened to take out the object.

View more quality glove box with low price and free shipping all over the world. 


  • Apr 09, 2021
  • Category: Glove Box
  • Comments: 0
Comments: 0

No comments

Leave a Reply

Your email address cannot be published. Required fields are marked*