1. Turn on the glove box
1) Turn on the main power switch (red);
2) Open the working gas pressure reducing valve (the main valve opens to a maximum of 15 seconds, and the auxiliary valve opens to 0.4~0.6Mpa);
3) Start the analyzer (the oxygen indicator starts to display when the water indicator drops below 200ppm, and the cycle can be opened only after the H2O & O2 is less than 200ppm, otherwise the cycle needs to be purged below 200ppm);
4) Start the cycle (H2O & O2 value <200ppm);
5) Start the chiller (please ignore if there is no);
6) Set the tank pressure to +1 ~ +6mbar.
2. Shut down the glove box
1) Turn off the H2O & O2 analyzer;
2) Close the loop;
3) Turn off the vacuum pump;
4) Chamber pressure supplement to 0mbar;
6) Close the main valve of the working gas pressure reducing valve;
7) Turn off the main power switch