Put the items in the glove box

 

1) Replenish air to the antechamber and close it after the replenishment is complete;

 

2) Open the outer antechamber door, put items in the antechamber, and close the outer antechamber door;

 

3) Air pumping and air replenishing three times each (LCD screen, airtight bottle, etc. cannot pump too much negative pressure, you can reduce the pumping time and increase the number of pumping as appropriate);

 

4) Open the inner antechamber door and take out the items;

 

5) Close the inner antechamber door, pump it to a negative pressure state, and place it for use.

 

Note:

 

It is forbidden to put absorbent articles such as rags, paper towels, etc. directly, and dry them before putting them in!

 

It is forbidden to try to open the inner and outer antechamber doors under negative pressure!

 

It is forbidden to carry objects that can damage the gloves, such as rings, watches, etc., when operating the gloves!

 

Take the items out of the glove box

 

1) Replenish air to the antechamber and close it after the replenishment is complete;

 

2) Open the inner antechamber door (Note: Before opening, make sure the atmosphere in the antechamber is pure, it is not sure, it needs to be pumped and supplemented three times to make it pure);

 

3) Put items into the antechamber and close the inner antechamber door;

 

4) Open the outer antechamber door and take out the items;

 


5) Close the outer antechamber, pump for three times, and finally pump to a negative pressure state, then place it for use.